KL-01
In vitro kidney/liver perfusion system
In vitro kidney/liver perfusion system
The modular design of the bath system is advantageous for quick installation and maintenance. The main units are illustrated above. No special technical background is required to assemble the units of the bath system or to replace them in case of failure. The replacement of individual units can be done at user level.
Perfusion solution, pressure measurement: T-joint connected to the CH2 chamber outlet stub with a silicone tube for a detachable connection.
Preparing bath system for measurement
TECHNICAL FEATURES | KL-01 |
---|---|
Application | In vitro kidney/liver perfusion system |
Testable organ types | Isolated kidney or liver from small rodents (rat, mouse, guinea pig) |
Physiological temperature humidity chamber | |
Fluid pressure does not stress the organ | |
Organ does not dry out | |
Organ is under circumstances similar to physiological conditions | |
The organs can be kept alive by continuous perfusion through the prepared arteries and veins | |
Perfusion pump circulates the perfusion fluid | |
Oxygenation of organs is ensured by a continuous flow of carbogen through the perfusion fluid | |
Temperature and vascular pressure can be continuously monitored | |
Isolated study of physiological and pathophysiological processes in the liver and kidney | |
Translational development of the method of liver and kidney transplantation | |
Isolated studies of drug and drug metabolism, bile excretion, renal and hepatic toxicity | |
Analysis of the interactions of individual active substances during hepatic and renal excretion | |
Structure of the organ to be examined is retained | |
Hormonal and neural factors can be excluded | |
Extremely high doses can be tested | |
Dimension | 200×100×150mm |
Weight | 1kg |
Thermostat | TC-02 External thermostat |
Peristaltic pump (Optional) | |
Pressure and temperature meter (Optional) | HC-B |
TECHNICAL FEATURES | TC-02 Thermostat |
---|---|
Application | Thermostat |
Temperature display | LED number display 3 digit |
Temperature display | Semiconductor |
Temperature control | Analogue PI control |
Temperature setting | Analogue mode continuous |
No. of channels | 2 independent thermostats |
Working temperature range | Room temperature … +40°C |
Temperature stability | ±0,01°C |
Power input | 250W |
Dimension | 320×350×220mm |
Weight | 10kg (without liquid) |
Power supply | On request: 110V or 220V, 50/60 Hz |
TECHNICAL FEATURES | HG-B |
---|---|
Application | Two-channel pressure and temperature meter |
HG Bandwidth | DC, 200Hz* |
HG Calibration levels | GND, 1V |
HG Bridge voltage | On demand |
HG Offset | ±1V (100Hgmm) relative to output |
HG Display | LED band |
HG Input impedance | 10MΩ symmetric |
HG Output impedance | 100Ω asymmetric |
HG Value conversion: | 100Hgmm = 1V |
TH Bandwidth | DC, 5Hz* |
TH Calibration levels | GND, 45°C (450mV) |
TH Bridge amplification | On demand |
TH Display | LED number display 3 digits |
TH Input impedance | 10kΩ symmetric |
TH Output impedance | 100Ω asymmetric |
TH Sensor type | PT1000 |
TH value conversion | 10°C = 0.1V |
Temperature accuracy | ±0,1°C |
Power supply | On request: 110V or 220V, 50/60 Hz |
*HG = pressure meter, TH temperature meter |